Index, Manual sample – Yokogawa Data Acquisition with PID Control CX2000 User Manual
Page 140
6-25
M-4659
Configuring the MV1000/MV2000
1
2
3
4
5
6
7
8
Index
Match Time Timer
Set the time match condition used in event action. These timers are also used in TLOG
computation of the computation function. You can set four match time timers (1 to 4).
• Kind
Daily Set the time match condition of a day.
Weekly Set the time match condition of a week.
Monthly Set the time match condition of a month.
Year Sets the time match condition for a year.
Set the items with check marks in the following table depending on the Kind setting.
Daily
Weekly
Monthly
Year
Month
Day
Week
Hour:Minute
Kind
Setup Item
• Month, Day, Week, Hour:Minute
Set the month, day, and weekday. Set the time in the range of 00:00 to 23:59 for
Hour:Minute.
• Timer action
Single Executes the action once when the condition is met.
Repeat Executes the action at every specified time.
Manual Sample
On a RD-MV2000 with the external input channel (/MC1) option, specify the channel that
will be manually sampled. On all other models, all channels will be manually sampled so
this setting is not necessary.
Click to display a channel selection dialog box.
Channel selection dialog box
Fill and increment by 1 starting from the top of the selection
Turn ON/OFF all in the selection at once
Drag to select a range
Manual sample number
Select a number from 001 to 120. The instantaneous values are output in this order.
Manual Sample
• Use
Select On when assigning a channel to the manual sample number.
• CH No.
Enter a channel number of a measurement channel, computation channel (/M1 and /
PM1 options), or external input channel (/MC1 option).
6.5 Entering General Settings