Scenario events – Teledyne LeCroy SierraFC M8-4 User Manual User Manual
Page 295

SierraFC M8‐4 Protocol Analyzer User Manual
295
Settings in Advanced Mode
Teledyne LeCroy
Scenario Events
A Scenario is a script you create using simple mouse clicks and text entries. As you work,
the script takes shape in the Scenario area of the application display. You can think of the
Scenario area itself as consisting of two subareas: A Global Rules area at the top, where
you create the Global Rules, and a Sequence area beneath the Global Rules, where you
create any sequences. Whether you are creating Arm Events, Jam Events or Jam Actions
(in Simple Mode), or Global Rules or a Sequence (in Advanced Mode), the menu‐driven
interface prompts you to specify the Event(s) for which you want to trigger Actions.
Figure 4.19: Event Properties Dialog
While many Events are line conditions, an Event also can be a condition that occurs within
a FC InFusion device (for example, detection of a trigger signal from another device).
Table 4.1 on page 267
and
Table 4.2 on page 274
lists supported Events in Simple Mode
and Advanced Mode respectively. Note that some Events are applicable only in the
context of creating sequences, in Advanced Mode (those Events appear on the drop‐
down list only if you are creating a sequence). Sequences can have multiple states, and
they allow branching between states.