Setting trigger conditions – Teledyne LeCroy SierraFC M8-4 User Manual User Manual
Page 172

Teledyne LeCroy
Advanced Mode (User‐Defined)
172
SierraFC M8‐4 Protocol Analyzer User Manual
Setting Trigger Conditions
To set the If and Else If trigger condition:
1. Click the Add Pattern button
for a Pattern field and choose a trigger
condition from the drop‐down list.
Figure 2.132: Choosing a Trigger Condition
2. Define each selected pattern in the same way as in Easy Mode, as described starting
. To use a timer, define it first.
Note:
You can set a timer for any If or Else If condition.
3. Enter a value for the number of occurrences before trigger in the Cont field, up to a
maximum of 65535 occurrences.
4. Choose a capture option: Everything, Nothing, or Pattern.
5. If you choose Pattern, you can select patterns for inclusion or exclusion. See
“Defining Patterns” on page 43
.
6. Choose pattern(s) and click the +>> button to add them for capture or exclusion. You
define each pattern the same way as in Easy mode (
7. For an output trigger, click the down arrow in the Ext. Out field and choose an
output trigger level. Note: Do not use the LOW setting in Advanced Mode.
8. To go to another state, click the down arrow in the Go To field and select a state. If
no other state has been defined, choose New State to add a state.