and total available memory to determine the actual number of samples or
segments, and time or points.
Setting up Sequence Mode (Adjacent)
1. Touch Timebase → Horizontal Setup... on the menu bar.
2. Click the Sequence tab.
3. Under Acquisition Settings, touch inside the Num Segments data
entry control and provide a value using your preferred input control
method. Additional information on using the touch screen controls
can be found in the Dialog Area
.
Note
: The number of segments you choose to display can be less than the
total number of segments in the waveform. For example, in the pop-up
images above, the number of display segments is 10, but the total number
of segments entered in the timebase dialog's
Num Segments
control is
100.
4. Touch the Enable Timeout checkbox.
5. Touch inside the Timeout data entry control and provide a timeout
value.
Note
: Use the sequence mode timeout to automatically interrupt the
sequence acquisition if the timeout value is exceeded without a valid
trigger. The timeout period accounts for instances when a
Num Segments
miscount occurs for some reason and the oscilloscope waits indefinitely for