Getting Started Manual

922136-00 Rev A

105

B

WaveScan allows searching through a single acquisition for a
measurement parameter that meets a variety of conditions, or
scanning through multiple acquisitions with allowable trigger actions
when conditions are met. In addition, a variety of views can be used
to understand the specific behaviors of searched or
scanned measurement parameters. See Introduction to WaveScan

(on

page 127

)

.

These capabilities are only possible with the measurement of every
occurrence of a measurement parameter in a long acquisition.

Both Track and Trend are tools that can be used to plot measurement data
and observe variations with respect to time. Differences between Track
and Trend are summarized in the following table:

Characteristic

Track

Trend

Representation

Parameter value vs.
time

Parameter value vs. event

Behavior

Non-cumulative
(resets after every
acquisition).
Unlimited number
of events

Cumulative over several
acquisitions up to 1 million
events

Time Correlation to
Other Data

Yes

No

Monitors an
Evolution in the
Frequency Domain

Yes

No. Trend points are not
evenly spaced in time and
therefore cannot be used for
an FFT.

Monitors the
Evolution of a
Measurement
Parameter over
Several Acquisitions

No. Track resets
after every
acquisition.

Yes

Ensures No Lost
Measurement Data

Yes. Maximum time
period that can be

No. Since data can be
accumulated over many