Reference, Specifications, Certifications – Teledyne LeCroy WaveRunner 6 Zi and 12-Bit HRO Getting Started Manual User Manual
Page 176: Emc compliance, Waverunner 6 zi and 12-bit hro
![background image](/manuals/353458/176/background.png)
WaveRunner 6 Zi and 12-Bit HRO
168
922136-00 Rev A
Reference
Specifications
Note
: Specifications are subject to change without notice.
Please refer to the Teledyne LeCroy website at
detailed specification information is regularly maintained on corresponding
product datasheets.
Certifications
This section contains the instrument’s Electromagnetic Compatibility
(EMC), Safety and Environmental certifications.
EMC Compliance
EC
D
ECLARATION
O
F
C
ONFORMITY
-
EMC
The oscilloscope meets intent of EC Directive 2004/108/EC for
Electromagnetic Compatibility. Compliance was demonstrated to the
following specifications as listed in the Official Journal of the European
Communities:
EN 61326-1:2006, EN 61326-2-1:2006 EMC requirements for electrical
equipment for measurement, control, and laboratory use.
1
Electromagnetic Emissions:
CISPR 11:2003, Radiated and Conducted Emissions Group 1, Class A
2 3
EN 61000-3-2:2006 Harmonic Current Emissions, Class A
EN 61000-3-3/A2:2005 Voltage Fluctuations and Flickers, Pst = 1
Electromagnetic Immunity:
EN 61000-4-2:2001 Electrostatic Discharge, 4 kV contact, 8 kV air, 4 kV
vertical/horizontal coupling planes
4
EN 61000-4-3:2006 RF Radiated Electromagnetic Field, 3 V/m, 80-1000
MHz; 3 V/m, 1400 MHz - 2 GHz; 1 V/m, 2 GHz - 2.7 GHz
4
EN 61000-4-4:2004 Electrical Fast Transient/Burst, 1 kV on power supply
lines, 0.5 kV on I/O signal data and control lines
4
EN 61000-4-5:2006 Power line Surge, 1 kV AC Mains, L-N, L-PE, N-PE
4
EN 61000-4-6:2007 RF Conducted Electromagnetic Field, 3 Vrms, 0.15 MHz
- 80 MHz
4