WaveRunner 6 Zi and 12-Bit HRO

120

922136-00 Rev A

Depending on your oscilloscope model, you can configure up to 8 pass/fail
conditions. The Boolean conditions to determine if your waveform passes
are as follows:

Pass/Fail Boolean Conditions

All True

All False

Any True

Any False

All Q1 to Q4

Or

All Q5 to Q8

Any Q1 to Q4

And

Any Q5 to Q8

P

ASS

/F

AIL

T

ESTING

S

ETUP

1. Touch Analysis → Pass/Fail Setup... on the menu bar.

2. Touch the Actions tab.

3. Touch the Enable Actions checkbox. This causes selected actions to

occur on your waveform's passing or failing a test.

4. Touch the Summary View checkbox to enable a line of text (as

follows) showing a concise status of the last waveform and keeping a
running count of how many sweeps have passed.

5. Touch inside the Pass If control, and select a Boolean condition from

the choices shown.

6. If you want to set up the test to end after a finite number of sweeps,

touch the Stop Test checkbox. Then touch inside the After data
entry control and provide a value using your preferred input control
method. Additional information on using the touch screen controls
can be found in the Dialog Area

(on

page 42

)

.

7. Use either the Pass or Fail button on the If section of the dialog to

set actions taking place when your waveform passes or fails the test.

8. The Then section of the dialog provides stop test, sound alarm, print

result, emit pulse, or save (the waveform) actions. If you want to
have the results printed (and your oscilloscope is not equipped with
a printer) make sure your instrument is connected to a local or
network printer. Learn more in the Printing and Hardcopy Functions
topic

(available in the online help files)

.