B

Comparing a Single Parameter

1. With the ParamCompare condition selected on a Qx dialog, touch

inside the Compare Values field and select All or Any from the pop-
up menu. By selecting All, the test are true only if every waveform
falls within the set limit. Selecting Any makes the test true if just one
waveform falls within the limit.

2. Touch inside the Condition field on the ParamCompare part of the

dialog and select a math operator from the pop-up menu.

3. Touch inside the Limit field and enter a value using the pop-up

numeric keypad. The value entered acquires the dimensions
corresponding to the parameters being tested.

For example, if you are testing a time parameter, the unit is seconds.

If you choose either WithinDeltaPct or WithinDeltaAbs from the
Condition menu (respectively, previous), you also have the choice of
setting the limit by means of the statistical buttons (relating to
standard statistical limits) at the bottom of the ParamCompare
dialog:

Comparing Dual Parameters

In Dual Parameter Compare mode, your oscilloscope can take parameter
results measured on two different waveforms and compare them to each
other. If desired, set your test to be true if Any or All waveforms fit the
criteria set by the comparison condition. Setups are conveniently shown in