Customizing menu steps – Fluke Biomedical medTester 5000C User Manual
Page 92

medTester 5000C
Operators Manual
5-22
4.
CYCLE TEST TIME IN MINUTES—Enter the time between
measurements (1-9 minutes).
5.
TRIGGER LEVEL—Enter the trigger threshold, that is the
percent of deviation in relative humidity that causes an automatic
measurement (1-9 percent).
Measurements of the relative humidity begin and display on the top line. Line
two of the display indicates the measurement interval and the trigger point
percentage.
Ctrl+P
triggers the readings. To terminate readings, press:
DONE
F5
End-of-Test Prompts
At the end of the autosequence, medTester prompts you for the following
data: (prompts can be turned off)
•
COMMENTS
•
NEXT TEST DUE DATE
•
USER TIME
•
FINISHED—STORE RECORD?
(appears if memory was selected)
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)