What™s a checklist – Fluke Biomedical medTester 5000C User Manual
Page 163

Biomedical Test System
IV Pump Module
10
10-25
4.
RATE LOW LIMIT: Enter the low limit for the flow rate. 0000.0
turns off low limit checking.
5.
TEST END TYPE: Select
6.
MANUAL to end the test by pressing a key.
7.
TIMED to end the test after a programmed time:
8.
END TIME: Enter the elapsed time that will end the
test, 00:01 to 99:59 (hours:minutes).
9.
VOLUME to end the test after a specific volume.
10.
END VOLUME: Enter the volume that will end the
test, 0000.1 to 9999.0 ml.
IPT-MC
For the IPT-MC, there is only 1 type of test available, a regular flow test:
1.
FLOW RATE: Enter the nominal flow rate, 0000.1 to 0999.0
ml/hr.
2.
RATE HIGH LIMIT: Enter the high limit for the flow rate.
0000.0 turns off high limit checking.
3.
RATE LOW LIMIT: Enter the low limit for the flow rate. 0000.0
turns off low limit checking.
4.
TEST END TYPE: Select
5.
MANUAL to end the test by pressing a key.
6.
TIMED to end the test after a programmed time:
7.
END TIME: Enter the elapsed time that will end the
test, 00:01 to 99:59 (hours:minutes).
8.
VOLUME to end the test after a specific volume.
9.
END VOLUME: Enter the volume that will end the
test, 0000.1 to 9999.0 ml.
After completing these entries, for up to ten (10) steps, the medTester
prompts you to store the autosequence:
STORE
F5
Print Your Autosequence
From the IV pump autosequences menu described above, printing any
autosequence is accomplished through this menu command: