Fluke Biomedical medTester 5000C User Manual
Page 238
medTester 5000C
Operators Manual
15-10
Each autosequence can contain up to ten SPO2 simulation steps, except the
INDEX 2XL which can contain up to 15 SPO2 steps. After those, several
special tests will be performed if they have been programmed into the
autosequence.
SPO2 Simulation Steps
For each step the medTester commands the SPO2 simulator to set up a
simulation of the programmed SPO2 in percentage, heart rate in beats per
minute (BPM), and amplitude in percentage. For a given step, only the
commands necessary for a change are sent.
For the Oxitest Plus, the medTester also sets up the programmed artifact.
For the CardioSat 100 and Index 2XL, the commands sent to the SPO2
simulator are noted on the medTester display to show you what is happening.
For the Index 2XL, the medTester can also be set up for Motion-Preset Tests
based on either Level 1 or Level 2 tests. The tables below show the eight
motion presets for each of the two levels of tests.
L
EVEL
1
P
RESET
N
AME
S
P
O2
% R
ATE
BPM
08 Normal/Tap
98
55
09 Normal/Shiver
98
55
10 Weak/Tap
90
95
11 Weak/Shiver
90
94
12 Brad/Tap
88
45
13 Hypoxic/Tap
70
95
14 Hypoxic/Shiver
70
95
15 Neonate/Shiver
90
180
L
EVEL
2
P
RESET
N
AME
S
P
O2
% R
ATE
BPM
16 BradyTap#2
88
45
17 Hypox/Tap#2
70
95
18 WeakTap#2
80
95
19 NormalTap#2
93
55
20 Asystole
91
90
21 LowFreq1
80
75
22 LowFreq2
70
75
23 SlowTap
80
75
For the CardioSat 100 and the INDEX 2XL, the commands sent to the
SPO2 simulator are noted on the medTester display to show you what is
happening.
The medTester waits for the programmed number of heartbeats for a settling
time for the pulse oximeter. The medTester beeps when settling is finished.