Fluke Biomedical medTester 5000C User Manual
Page 272
medTester 5000C
Operators Manual
16-16
Begin by accessing the custom, autosequence, pacer menu as described earlier.
This time, select:
MAKE
F1
Next select the name of the autosequence you want to customize or create.
Follow the following steps, providing the desired information:
1.
Name—This is the name you want to give to the autosequence,
with a maximum of six characters.
2.
Menu Steps:
•
Load: Select internal 50 ohms or a TQA module
•
Pulse tests, 20 steps:
Current: 10 to 250 mA
Pulse Rate: 30 to 200 PPM
Load, if using a multiple test load TQA module
•
Current +/- limit: 0 to 99% (0% is off)
•
Pulse Rate +/- limit: 0 to 99 PPM (0 is off)
•
Demand Test:
Pacer Rate: 30 to 200 PPM
Underdrive ECG Rate: 30, 60, 80, 120, or 160 BPM
Overdrive ECG Rate: 60, 80, 120, 160, or 200 BPM
•
Async Test:
Pacer Rate: 30 to 200 PPM
Overdrive ECG Rate: 60, 80, 120, 160, or 200 BPM
•
Refractory Period Test
After customizing your autosequence, the medTester prompts you to store
the autosequence:
STORE
F5
Print Your Autosequence
From the pacer custom autosequences menu described above, printing any
autosequence is accomplished through this menu command:
F3