For cardiosat 100 – Fluke Biomedical medTester 5000C User Manual
Page 245
Biomedical Test System
SPO2 Module
15
15-17
For CardioSat 100:
SPO2 Autosequence Number: A72
Autosequence Name: HPMRLN
Manufacturer: Hewlett Packard Merlin Series
Settling Time: 25 Beats
Heart Rate Limit:
± 4 BPM (beats per minute)
Step Number
SPO2
(in percent)
SPO2 Limit
(in percent)
Heart Rate
(in BPM)
Amplitude
(in percent)
1 96
± 3
75 100
2 98
± 3
230 100
3 92
± 3
75 100
4 82
± 4
75 100
5 72
± 5
75 100
6 98
± 3
45 100
Starting Alarm Value for SPO2: 96%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 100%
SPO2 Alarm Test: Hi–100% Lo–85%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
Probe Test
SPO2 Autosequence Number: A73
Autosequence Name: DATEX
Manufacturer: Datex
Settling Time: 25 Beats
Heart Rate Limit:
± 4 BPM (beats per minute)
Step Number
SPO2
(in percent)
SPO2 Limit
(in percent)
Heart Rate
(in BPM)
Amplitude
(in percent)
1 96
± 3
75 100
2 98
± 3
230 100
3 92
± 3
75 100
4 82
± 4
75 100
5 72
± 5
75 100
6 98
± 3
45 100
Starting Alarm Value for SPO2: 96%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 100%
SPO2 Alarm Test: Hi–100% Lo–85%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
Probe Test