Standard Safety Autosequence Names
You can access autosequences from either medTester menu commands or
from the keyboard shortcut commands. See the list of autosequences below.
SAFETY AUTOSEQUENCES
Autosequence
Name
Function
Menu Command
From medTester 5000C Main Menu
(MENU 1)
Keyboard
Shortcut
Commands
10LD X
10 Lead
External Test
AUTO
F1
10LDX
F1
SAFETY
F1
A1
3LD X
3 Lead External
Test
AUTO
F1
3LDX
F2
SAFETY
F1
A2
4LD X
4 Lead External
Test
AUTO
F1
4LDX
F3
SAFETY
F1
A3
CASE X
Case External
Critical Device,
No Leads
CASE X
F4
SAFETY
F1
AUTO
F1
A4
GEN X
General Device,
No Leads
GEN X
F5
SAFETY
F1
AUTO
F1
A5
From MENU 2
10LD I
10-Lead
Internal Test
AUTO
F1
10LD I
F1
SAFETY
F1
A6
3 LD I
3-Lead Internal
Test
AUTO
F1
3LDI
F2
SAFETY
F1
A7
4 LD I
4-Lead Internal
Test
AUTO
F1
4LDI
F3
SAFETY
F1
A8
CASE I
Case Internal
Critical Device,
No Leads
CASE I
F4
SAFETY
F1
AUTO
F1
A9
GEN I
General Device
Internal, No
Leads
GEN I
F5
SAFETY
F1
AUTO
F1
A10
From MENU 3
blank User-defined
AUTO
F1
blank
F1-F5
SAFETY
F1
A11-A15
From medTester 5000C Main Menu
LINEMON Line
Voltage
Monitor
AUTO
F1
LINEMON
F4
A16
ENVMON Environment
Monitor
AUTO
F1
ENVMON
F5
A17