Fluke Biomedical medTester 5000C User Manual
Page 288

medTester 5000C
Operators Manual
17-14
After customizing your autosequence, the medTester prompts you to store
the autosequence:
STORE
F5
Print Your Autosequence
From the NIBP custom autosequences menu described above, printing any
autosequence is accomplished through this menu command:
F3
Reinitialize Factory Default Settings
Any customized autosequences can be returned to the factory default settings.
To do this, enter these medTester commands:
1.
From the medTester 5000C main menu,
MENU
1, press the
right arrow to access
MENU2
.
2.
Enter the command:
CUSTOM
F3
3.
From the custom menu, press the right arrow and enter the
command:
INIT
F3
4.
Press the right arrow, and then press:
NIBP
F2
5.
To initialize NIBP autosequences to defaults, press:
YES
F5
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)