Fluke Biomedical medTester 5000C User Manual
Page 227

Biomedical Test System
ESU Module
14
14-17
•
Expected Power: Enter the power expected to be generated
by the ESU, from 0 to 500 W.
•
Power high limit (00% = off)
•
Power low limit (00% = off)
RF Leakage Test Type
•
Electrode: Select either Active or Dispersive electrodes to
test.
•
Load Status: Select Open or Closed. For RF303
RS
, use only
Open Load Status.
•
Load (if Load Status Closed): Select an internal load (same as
for Generator Output).
•
Mode: Same as for Generator Output
•
Output: Same as for Generator Output
•
Power Setting: Same as for Generator Output
Contact Quality Monitor (CQM) Test Type
Nothing to select.
Return Fault Test Type (not applicable to RF303
RS
)
•
Test Side: Select either Cut- or Coag-side ESU mode
•
Power Setting: Same as for Generator Output
After each step selection, you can press
MORE
F4
to select more steps, or
END
F5
to end your selections.
After customizing your autosequence, the medTester prompts you to store
the autosequence:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)