Fluke Biomedical medTester 5000C User Manual
Page 287
Biomedical Test System
Non-Invasive Blood Pressure (NIBP) Module
17
17-13
Begin by accessing the custom, autosequence, NIBP menu as described
earlier. This time, select:
MAKE
F1
Next select the name of the autosequence you want to customize or create.
Follow the following steps, providing the desired information:
1.
Name—This is the name you want to give to the autosequence,
with a maximum of six characters.
2.
Menu Steps:
•
NIBP Tests, 10 steps:
NIBP
selection:
60/ 30( 40) ADULT
80/ 50( 62) ADULT
100/ 65( 75) ADULT
120/ 80( 90) ADULT
150/100(115)
ADULT
200/150(165)
ADULT
255/195(215)
NEONATE
60/ 30( 40) NEONATE
80/ 50( 62) NEONATE
100/ 65( 75) NEONATE
120/ 80( 90) NEONATE
150/100(115)
NEONATE
120/ 80( 90) PAC
120/ 80( 90) PVC
120/ 80( 90) AFIB
120/ 80( 90) MBC
120/ 80( 90) ASC
Heart Rate: (except for arrhythmias)
30 to 240 BPM (80 nominal)
Envelope Gain: (except for arrhythmias)
0 to 200 % (100 nominal)
Envelope Shift: (except for arrhythmias)
Zero
Positive: 0 to 100
Negative: 0 to 100
•
NIBP +/- LIMIT: 0 t0 99 % (0 off, 20 nominal)
•
HEART RATE +/- LIMIT: 0 to 99 BPM
(0 off, 4 nominal)
•
PRESSURE TEST
•
LEAK TEST
•
POP OFF TEST