Leakage to groundšall leads, Leakage to groundšindividual lead, Manual tests – Fluke Biomedical medTester 5000C User Manual
Page 47: Chapter 4, Performing manual tests

4-1
Chapter 4
Manual Tests
Learn to use the medTester 5000C to manually perform 
electrical safety tests for line voltage, leakage current, 
equipment current, and resistance. 
Performing Manual Tests
The medTester 5000c has all the features of a manual safety tester. You can 
access these features from the items under 
MANUAL
on Menu 1.
Printing Manual Measurements 
With the RS-232/Printer module installed, you can print any manual 
measurement result. Results print out on a single line. To print manual 
measurements with the measurement displayed on the medTester, 
press : 
Ctrl+P.
The current measurement prints through the port that you activate from the
UTIL
menu:
OUTPUT
F2
UTIL
F5
RECORD
F1
Line Voltage 
See Figure 4-1, Three Types of Line Voltage Measurements. 
Three line voltage measurements are available within a range of 0-200 volts 
RMS with an accuracy of 
±5% of the range.