Comments, Next test due date, User time – Fluke Biomedical medTester 5000C User Manual
Page 285: User time is required, Print test tag, Store or print test record

Biomedical Test System
Non-Invasive Blood Pressure (NIBP) Module
17
17-11
NIBP Autosequence Number: A83
Autosequence Name: NEONT1
Step Number
Blood Pressure
Heart Rate in BPM
(beats per minute)
Envelope Gain
Envelope Shift
1
80/50 (62) NEONT
120
100%
+0
2
60/30 (40) NEONT
80
100%
+0
3
100/65 (75) NEONT
180
100%
+0
NIBP Test Limits:
• NIBP: ± 20%
• Heart Rate: ± 6 BPM
Pressure Test: No
Leak Test: No
Pop Off Test: No
NIBP Autosequence Number: A84
Autosequence Name: NEONT2
Step Number
Blood Pressure
Heart Rate in BPM
(beats per minute)
Envelope Gain
Envelope Shift
1
80/50 (62) NEONT
120
100%
+0
2
60/30 (40) NEONT
80
100%
+0
3
100/65 (75) NEONT
180
100%
+0
NIBP Test Limits:
• NIBP: ± 20%
• Heart Rate: ± 6 BPM
Pressure Test: Yes
Leak Test: Yes
Pop Off Test: Yes
Posttest Prompts
When all tests are completed, the medTester 5000C prompts you for:
1.
COMMENTS
2.
NEXT TEST DUE DATE
3.
USER TIME
EMS/CMMS system users:
User time is required.
4.
PRINT TEST TAG
5.
STORE or PRINT TEST RECORD