Esu autosequence descriptions – Fluke Biomedical medTester 5000C User Manual
Page 216

medTester 5000C 
Operators Manual 
14-6
To accept previously input prompt information, press
YES
, and go to the Test
Sequence section. If you choose
NO
, go to the Pretest Device Prompts section that
follows.
Pretest Device Prompts
There are eight prompts for information that can attach to test records. After 
selecting the desired autosequence, the medTester 5000C prompts you to 
enter device prompts. If you have some or all prompts turned off, the 
autosequence still gives you the brief opportunity to enter device information 
at the tests initiation with the message: 
1.
OP CODE—This is the operator code for the person doing the 
testing. This code can consist of a maximum of three characters 
which can be entered from the medTester keyboard. An example 
of an operator code could be a person’s initials or employee 
number. The previous operator’s code, if any, will appear. If the 
current operator has not changed from previous use, press the 
ENT
key.
2.
TYPE—Enter the device name, code, or other descriptor for the 
equipment under test. This entry can be a maximum of 16 
characters. 
3.
MANF—Enter the name of the device manufacturer, 16 
characters maximum. 
4.
LOC—Type in the location of the device, 16 characters 
maximum. The previous location, if any, will appear. If this occurs 
and is the desired location, press the 
ENT
key.
5.
MODEL—Enter the device model number, 16 characters 
maximum. 
6.
SERIAL #—Type the device serial number, again 16 characters 
maximum. 
7.
CONTROL #—Enter a control number with a maximum of 16 
characters. 
Press F5 for all device info prompts
