Customizing iv pump tests, Make your own autosequence, Making an autosequence – Fluke Biomedical medTester 5000C User Manual
Page 156

medTester 5000C 
Operators Manual 
10-18
Pressure Test (Infutest, IPT-1, IPT-MC and IDA 4 Plus) 
• After completing the flow test, the autosequence continues by offering to 
perform a pressure test. The pressure test measures the peak occlusion 
pressure and the elapsed time to reach the peak. Press F1 YES to run the 
pressure test or F2 NO to skip it and go to the next flow test step. 
• The pressure test runs the same for all IV pump analyzers. The IV pump
analyzer must be primed in the same manner as for the flow test. Then 
press F5 START to start the pressure test. 
• The IV pump analyzer occludes (blocks) the IV line, measures the
pressure developed and notifies the medTester. The medTester displays 
the pressure (PRESSURE=) in psi (pounds per square inch) and mmhg 
(millimeters of mercury). The medTester keeps track of the peak pressure, 
which is the highest pressure encountered during the test, and displays it 
(PK=) also in PSI and mmHg. The medTester also displays the elapsed 
time (ET=). 
• Allow the test to run long enough for the IV pump to sense the
occlusion, alarm, and to shut off, and for the pressure to decrease below 
the peak. Press 
F5
END to finish the pressure test.
• The final display gives you the peak pressure level. The
@
sign (with a
time reading) indicates the time when the peak pressure was reached.
• You can repeat the pressure test by pressing
F4
. To continue with the next
step in this autosequence, press
F5
.
Note
The remaining steps in this and other IV pump autosequences 
follow the same sequence described here. Consult the table IV 
Pump Autosequence Tests to see the steps in each autosequence. At 
the end of the last step, the medTester 5000C proceeds to the 
posttest prompts. 
Posttest Prompts 
When all tests are completed with all Posttest prompts turned on, the 
medTester 5000C prompts you for: 
1.
COMMENTS
2.
NEXT TEST DUE DATE
3.
USER TIME
Computerized Maintenance Management
Software (CMMS) users:
User time is required.
4.
PRINT TEST TAG?
5.
STORE and/or PRINT TEST RECORD