Posttest prompts, Customizing spo2 autosequences, Make your own autosequence – Fluke Biomedical medTester 5000C User Manual
Page 246: For cardiosat 100

medTester 5000C
Operators Manual
15-18
For CardioSat 100:
SPO2 Autosequence Number: A74
Autosequence Name: GENERC
Manufacturer: Generic–Not Specified
Settling Time: 25 Beats
Heart Rate Limit:
± 4 BPM (beats per minute)
Step Number
SPO2
(in percent)
SPO2 Limit
(in percent)
Heart Rate
(in BPM)
Amplitude
(in percent)
1 99
± 4
250 100
2 96
± 4
230 100
3 92
± 4
205 100
4 88
± 5
180 100
5 78
± 5
155 100
6 75
± 5
130 100
7 65
± 6
105 100
8 55
± 6
80 100
9 45
± 6
55 100
10 35
± 6
30 100
11 90
± 4
95 100
12 88
± 5
45 100
13 85
± 5
130 20
14 92
± 4
95 40
15 93
± 4
90 50
Starting Alarm Value for SPO2: 96%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 100%
SPO2 Alarm Test: Hi–100% Lo–85%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)