Customize your medtester, Toolbox monitors – Fluke Biomedical medTester 5000C User Manual
Page 89

Biomedical Test System
Autosequences
5
5-19
End-of-Test Prompts
At the end of the autosequence, medTester prompts you for the following
data: (prompts can be turned off)
•
COMMENTS
•
NEXT TEST DUE DATE
•
USER TIME
.
•
STORE RECORD?
(if memory was selected), or
PRINT?
(if print was
selected
Toolbox Monitors
The toolbox menu on the medTester 5000C main menu contains monitoring
sequences for temperature and relative humidity. To add these two options to
your medTester, contact Extech Instruments Coporation.
Temperature Monitor
You can monitor temperature readings over periods of time, print the results,
and store the results in memory. To access the monitor from the keyboard,
enter:
TEMPMON
F4
TOOLBOX
F2
Selections then display as:
Select
MEMORY
if you want to store the test record. This requires both the RS-
232/Printer module and the 100 Record Memory module and . Select
if
you want to print the test record. Select
BOTH
if you want to store and print
the record. Follow these set-up instructions:
1.
Plug the temperature adapter test leads into the
EXT INPUT
jacks
on medTester’s front panel.
2.
Set the adapter’s power switch to the desired temperature scale,
Fahrenheit or Celsius. It must match the medTester selection.
3.
Press the
°F/°C (F2)
key for the desired temperature scale.
RUN TEMPERATURE MONITOR A18 TO:
MEMORY
F1
F2
NEITHER
F4
BOTH
F3