For the oxitest plus – Fluke Biomedical medTester 5000C User Manual
Page 242
medTester 5000C
Operators Manual
15-14
For the Oxitest PLUS:
SPO2 Autosequence Number: A72
Autosequence Name: ADULT3
Settling Time: 25 Beats
Heart Rate Limit:
± 4 BPM (beats per minute)
Step
Number
SPO2
(in percent)
SPO2 Limit
(in percent)
Heart Rate
(in BPM)
Amplitude
(in percent)
Artifact
1 97 ± 4
75 100
Off
2 97 ± 4
45 100
Off
3 97 ± 4
180 100
Off
4 80 ± 4
45 100
Off
5 90 ± 4
120 100
Off
6 80 ± 4
180 100
Off
7 97 ± 4
66
4
Off
8 97 ± 4
70 33
Movement
9 97 ± 4
80 52
Tremor
Starting Alarm Value for SPO2: 93%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 100%
SPO2 Alarm Test: Hi–97% Lo–80%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
Probe Test
SPO2 Autosequence Number: A73
Autosequence Name: NEO1
Settling Time: 25 Beats
Heart Rate Limit:
± 6 BPM (beats per minute)
Step
Number
SPO2
(in percent)
SPO2 Limit
(in percent)
Heart Rate
(in BPM)
Amplitude
(in percent)
Artifact
1 93 ± 4
145
40
Off
2 93 ± 4
75
40
Off
3 93 ± 4
225
40
Off
4 97 ± 4
225
40
Off
5 80 ± 4
60
40
Off
6 93 ± 4
145
40
Movement
7 93 ± 4
145
40
Tremor