Isolation Lead Leakage
See Figure 4-7, ECG Lead Isolation Test: All Leads and Individual Leads.
You can do the following two tests in an all-lead or individual-lead mode. The
selected lead or leads are connected to the AAMI load with line voltage
applied to the lead(s). The test receptacle is fixed with normal polarity, closed
ground, and closed neutral during these tests.
WARNING!
During any isolation lead leakage test, 120 V at 1 mA is present
on the ECG leads. This is an electrical shock hazard.
Lead Isolation—All Leads
1.
Plug the EUT into the
TEST RECEPTACLE.
2.
Connect the ECG leads to be tested to their respective, color-
coded binding posts on the medTester 5000C.
3.
Enter menu commands:
µA mV
F2
MANUAL
F2
ECG
F4
ALL/ISO
F4
3.
To begin the measurement, press
ISOV ON
F5
. The test begins,
and the
ISO VOLTS
annunciator lights. The reading is displayed for
up to 15 seconds. You can escape to terminate the test. The test
self-terminates after 15 seconds.
Lead Isolation—Single Lead
1.
Plug the EUT into the
TEST RECEPTACLE.
2.
Connect the ECG lead to be tested to its respective color-coded
binding post on the medTester 5000C.
3.
Enter menu commands:
µA mV
F2
MANUAL
F2
ECG
F4
LD/ISO
F5