Posttest prompts, Esu autosequence descriptions – Fluke Biomedical medTester 5000C User Manual
Page 220

medTester 5000C
Operators Manual
14-10
ESU Autosequence Descriptions
Autosequence Number: A60
Autosequence Name: ASPENX
Step and Test
Type
Load
ESU Mode
Output
Power
Setting
Expected Power
or
Electrode
Limits
1
Generator Output
300 Ohms
Monopolar
Pure Cut
120
120 W
± 15%
2
Generator Output
300 Ohms
Monopolar
Pure Cut
300
300 W
± 15%
3
Generator Output
300 Ohms
Monopolar
Blend 1
69
69 W
± 15%
4
Generator Output
300 Ohms
Monopolar
Blend 1
180
180 W
± 15%
5
Generator Output
300 Ohms
Monocoag
Standard
48
48 W
± 15%
6
Generator Output
300 Ohms
Monocoag
Standard
120
120 W
± 15%
7
Generator Output
300 Ohms
Monocoag
Spray
40
40 W
± 15%
8
Generator Output
300 Ohms
Monocoag
Spray
80
80 W
± 15%
9
RF Leakage
Open Monocoag Standard 120 Active
Electrode N/A
10
RF Leakage
Open Monocoag Standard 120
Dispersive
Electrode
N/A
11
RF Leakage
Open
Monopolar
Pure Cut
300
Active Electrode
N/A
12
RF Leakage
Open Monopolar Pure
Cut 300
Dispersive
Electrode
N/A
13
Generator Output
50 Ohms
Bipolar
Coag
20
20 W
± 20%
14
Generator Output
50 Ohms
Bipolar
Coag
50
50 W
± 15%
15
Generator Output
50 Ohms
Bipolar
Cut
20
20 W
± 20%
16
Generator Output
50 Ohms
Bipolar
Cut
50
50 W
± 15%