Customizing spo2 autosequences – Fluke Biomedical medTester 5000C User Manual
Page 250

medTester 5000C 
Operators Manual 
15-22
Posttest Prompts
When all tests are completed, the medTester 5000C prompts you for:
1.
COMMENTS
2.
NEXT TEST DUE DATE
3.
USER TIME
EMS/CMMS system users:
User time is required
4.
PRINT TEST TAG
5.
STORE or PRINT TEST RECORD
Customizing SPO2 Autosequences
(For detailed information about customizing autosequences, please refer to 
Chapter 6, Customize Your medTester 5000C.) 
To begin customizing SPO2 autosequences, you can view any of your current 
ones by entering these menu commands: 
1.
From the main menu,
MENU1
, press the right arrow to access
MENU2.
2.
Enter the menu commands:
SPO2
F5
AUTOSEQ
F1
CUSTOM
F3
3.
From the SPO2
menu, select:
VIEW
F2
4.
Now select the autosequence you want to view.
5.
Finally scroll through the steps in the sequence by using your
UP
and
DOWN
arrow keys.
Make Your Own Autosequence 
You can customize any of the ten SPO2 autosequences, either the factory 
preprogrammed or the blank ones. This section describes the menu steps that 
you use to customize, or create, autosequences. Note that each customized 
autosequence can include as many as ten SPO2 simulation steps. After you 
enter the step parameters you want, the medTester prompts you for more 
steps, if desired. If you do not want further steps in your autosequence, you 
can press 
END
. Then you can program to include several special tests as
desired.