Remote commands list – Fluke Biomedical medTester 5000C User Manual
Page 186

medTester 5000C
Operators Manual
12-8
UNIT of MEASUREMENT ABBREVIATIONS
Abbreviation
Unit of Measurement
V RMS
Volts RMS
A RMS
Amperes RMS
µA RMS
Microamperes RMS
µA DC
Microamperes DC
OHMS Ohms
RPM
Revolutions per minute
DEG F
Degrees Fahrenheit
DEG C
Degrees Centigrade
% Percent
Error Responses
If you issue an incorrect or nonexistent remote command, the medTester
responds with an error statement in this format:
ERR=
xx. Here is a list of
possible error responses and explanations.
ERROR RESPONSES
Error Statement
Explanation
ERR=01,UNKNOWN COMMAND
The received command was not
recognized.
ERR=02,ILLEGAL COMMAND
The received command was not
allowed for the current instrument
status.
ERR=03,ILLEGAL PARAMETER
A parameter sent with a command
was not the right type, was out of
range, or was not understood, or
not the right number of parameters
were sent.
ERR=04,MODULE NOT INSTALLED
Can not execute the command
because the required instrument
module is not installed.
ERR=20,CURRENT SOURCE NOT
CONNECTED
For a resistance measurement
command, the current source
connection has not been made.
ERR=21,MEMORY FULL
For a file down transfer command,
the medTester memory is full and
can not accept anymore data to
fulfill the transfer.
ERR=22,GFCI
A ground fault circuit interruption
was triggered caused by a fault in
the EUT plugged into the test
receptacle. This error is not a
response to a command. It
appears whenever a ground fault
occurs.
ERR=23,WRONG FILE
For a file transfer command where