Remote commands list – Fluke Biomedical medTester 5000C User Manual
Page 186

medTester 5000C 
Operators Manual 
12-8
UNIT of MEASUREMENT ABBREVIATIONS
Abbreviation
Unit of Measurement
V RMS
Volts RMS
A RMS
Amperes RMS
µA RMS
Microamperes RMS
µA DC
Microamperes DC
OHMS Ohms 
RPM 
Revolutions per minute
DEG F
Degrees Fahrenheit
DEG C
Degrees Centigrade
% Percent
Error Responses
If you issue an incorrect or nonexistent remote command, the medTester 
responds with an error statement in this format: 
ERR=
xx. Here is a list of
possible error responses and explanations.
ERROR RESPONSES
Error Statement
Explanation
ERR=01,UNKNOWN COMMAND
The received command was not 
recognized. 
ERR=02,ILLEGAL COMMAND
The received command was not 
allowed for the current instrument 
status. 
ERR=03,ILLEGAL PARAMETER
A parameter sent with a command 
was not the right type, was out of 
range, or was not understood, or 
not the right number of parameters 
were sent. 
ERR=04,MODULE NOT INSTALLED
Can not execute the command 
because the required instrument 
module is not installed. 
ERR=20,CURRENT SOURCE NOT 
CONNECTED 
For a resistance measurement 
command, the current source 
connection has not been made. 
ERR=21,MEMORY FULL
For a file down transfer command, 
the medTester memory is full and 
can not accept anymore data to 
fulfill the transfer. 
ERR=22,GFCI
A ground fault circuit interruption 
was triggered caused by a fault in 
the EUT plugged into the test 
receptacle. This error is not a 
response to a command. It 
appears whenever a ground fault 
occurs. 
ERR=23,WRONG FILE
For a file transfer command where