Waveform groups – Fluke Biomedical medTester 5000C User Manual
Page 110

medTester 5000C
Operators Manual
7-2
Waveform Groups
You can generate any of the waveforms listed in the five groups in the table
below for the testing of patient monitors. The table’s first column lists the
names of the waveforms as they appear in the medTester 5000C display.
Column two describes the display names, and column three shows the remote
command for each waveform.
WAVEFORM GROUPS
medTester 5000C
Display Wave Menu
Name
Description
Remote Command
Group Name:
PERF
2Hz SQ
2 Hz Square
W001
4 Sec
4 Second Pulse
W002
10Hz
10 Hz Sine Wave
W003
40Hz
40 Hz Sine Wave
W004
60Hz
60 Hz Sine Wave
W005
100Hz
100 Hz Sine Wave
W006
1KHz
1 kHz Square Wave
W007
1HzTRI
2 Hz Triangle
W008
CMRR
60 Hz Common Mode Rejection Test
W009
.5Hz
0.5 Hz Sine Wave
W014
Group Name: ECG
30BPM
Normal Sinus ECG at 30 Beats Per
Minute (BPM)
W010
60BPM
Normal Sinus ECG at 60 BPM
W011
120BPM
Normal Sinus ECG at 120 BPM
W012
240BPM
Normal Sinus ECG at 240 BPM
W013
Group Name: ARRHY1
AFIB
Atrial Fibrillation 1, Course
W015
2°BLK1
2nd Degree A-V Block, Type 1,
Wenckebach
W016
PAC
Premature Atrial Contractions
W017
MB80
Missed Beat at 80 BPM
W018
MB120
Missed Beat at 120 BPM
W019
Group Name: ARRHY2
PVC1
PVC Type 1, Left
W020
PVC2
PVC Type 2, Right
W021
MFPVC Multi-focal
PVCs
W022
PVCRT
PVC 1, R on T
W023