Humidity monitor – Fluke Biomedical medTester 5000C User Manual
Page 87

Biomedical Test System
Autosequences
5
5-17
1.
OP CODE—The operator is the person doing the testing. This 
code can consist of a maximum of three characters which can be 
entered from the medTester keyboard or barcode Scan Gun. An 
example of an operator code could be a person’s initials or 
employee number. The previous operator’s code, if any, will 
appear. If the current operator has not changed from previous 
use, press the 
ENT
key.
2.
LOC—Type in the location of the device, 16 characters 
maximum. The previous location code, if any, will appear. If this 
occurs and is the desired location, press the 
ENT
key.
3.
CONTROL #—Enter a control number with a maximum of 16 
characters. 
4.
CYCLE TEST TIME IN MINUTES—Enter the time between 
measurements (1-9 minutes). 
5.
TRIGGER LEVEL—Enter the trigger threshold, that is the 
amount of deviation in hot-to-neutral voltage that causes an 
automatic measurement (1-9 volts). 
Test measurements of voltage begin and display. Every cycle or trigger 
measures and records hot-to-neutral voltage, neutral-to-ground voltage, and 
hot-to-ground voltage measurements. 
Ctrl+P
triggers the readings. When you
finish, press:
DONE
F5
End-of-Test Prompts
At the end of the autosequence, medTester prompts you for the following 
data: (prompts can be turned off) 
•
COMMENTS
•
NEXT TEST DUE DATE
•
USER TIME
•
FINISHED—STORE RECORD?
(appears if memory was selected)
Environmental Monitor 
This test lets you make ground potential and resistance measurements 
between a common ground point and up to 99 other ground points in a 
room. 