Humidity monitor – Fluke Biomedical medTester 5000C User Manual
Page 87

Biomedical Test System
Autosequences
5
5-17
1.
OP CODE—The operator is the person doing the testing. This
code can consist of a maximum of three characters which can be
entered from the medTester keyboard or barcode Scan Gun. An
example of an operator code could be a person’s initials or
employee number. The previous operator’s code, if any, will
appear. If the current operator has not changed from previous
use, press the
ENT
key.
2.
LOC—Type in the location of the device, 16 characters
maximum. The previous location code, if any, will appear. If this
occurs and is the desired location, press the
ENT
key.
3.
CONTROL #—Enter a control number with a maximum of 16
characters.
4.
CYCLE TEST TIME IN MINUTES—Enter the time between
measurements (1-9 minutes).
5.
TRIGGER LEVEL—Enter the trigger threshold, that is the
amount of deviation in hot-to-neutral voltage that causes an
automatic measurement (1-9 volts).
Test measurements of voltage begin and display. Every cycle or trigger
measures and records hot-to-neutral voltage, neutral-to-ground voltage, and
hot-to-ground voltage measurements.
Ctrl+P
triggers the readings. When you
finish, press:
DONE
F5
End-of-Test Prompts
At the end of the autosequence, medTester prompts you for the following
data: (prompts can be turned off)
•
COMMENTS
•
NEXT TEST DUE DATE
•
USER TIME
•
FINISHED—STORE RECORD?
(appears if memory was selected)
Environmental Monitor
This test lets you make ground potential and resistance measurements
between a common ground point and up to 99 other ground points in a
room.