Toolbox, Test receptacle, Csa label – Fluke Biomedical medTester 5000C User Manual
Page 22

medTester 5000C
Operators Manual
1-10
medTester 5000C Modules
medTester
5000C
Module Name
Functions Enabled
medTester
Prerequisites for Use of
Module
Required for...
SPO2
• SPO2
autosequences.
• RS-232/Printer Module.
• Memory—100 Records
Module.
• Expanded Memory
Module.
• Data Transfer Module.
• 10 autosequences for testing
SPO2 monitors with Fluke
Biomedical analyzers.
• Including SPO2
autosequences in checklists.
Pacer
• Transcutaneous
Pacemaker
autosequences.
• RS-232/Printer Module.
• Memory—100 Records
Module.
• Expanded Memory
Module.
• Data Transfer Module.
• 10 autosequences for testing
transcutaneous pacemakers
with the Fluke Biomedical
Impulse 4000.
• Including Pacer
autosequences in checklists.
NIBP
• Non-Invasive Blood
Pressure
autosequences.
• RS-232/Printer Module.
• Memory—100 Records
Module.
• Expanded Memory
Module.
• Data Transfer Module.
• 10 autosequences for testing
Non-Invasive Blood Pressure
monitors with the Fluke
Biomedical CuffLink.
• Including NIBP pump
autosequences in checklists.
Bar Code
(hardware)
• Scanning of bar
code data into test
record.
• None
• Scanning of data.