Toolbox, Test receptacle, Csa label – Fluke Biomedical medTester 5000C User Manual
Page 22

medTester 5000C 
Operators Manual 
1-10
medTester 5000C Modules
medTester
5000C
Module Name
Functions Enabled
medTester
Prerequisites for Use of
Module
Required for...
SPO2
• SPO2
autosequences.
• RS-232/Printer Module. 
• Memory—100 Records 
Module.
• Expanded Memory
Module.
• Data Transfer Module.
• 10 autosequences for testing
SPO2 monitors with Fluke 
Biomedical analyzers. 
• Including SPO2
autosequences in checklists.
Pacer
• Transcutaneous
Pacemaker 
autosequences. 
• RS-232/Printer Module. 
• Memory—100 Records 
Module.
• Expanded Memory
Module.
• Data Transfer Module.
• 10 autosequences for testing
transcutaneous pacemakers 
with the Fluke Biomedical 
Impulse 4000. 
• Including Pacer
autosequences in checklists.
NIBP
• Non-Invasive Blood
Pressure 
autosequences. 
• RS-232/Printer Module. 
• Memory—100 Records 
Module.
• Expanded Memory
Module.
• Data Transfer Module.
• 10 autosequences for testing
Non-Invasive Blood Pressure 
monitors with the Fluke 
Biomedical CuffLink. 
• Including NIBP pump
autosequences in checklists.
Bar Code 
(hardware) 
• Scanning of bar
code data into test 
record. 
• None
• Scanning of data.
