Print your autosequence, Reinitialize factory default settings – Fluke Biomedical medTester 5000C User Manual
Page 251

Biomedical Test System
SPO2 Module
15
15-23
Note
You can find detailed information about customizing
autosequences in Chapter 6, Customize Your medTester 5000C.
Configuring for Your SPO2 Simulator
The medTester can only be configured for one type of SPO2 Simulator at a
time, either the Oxitest PLUS (including Oxitest PLUS 7), the CardioSat 100
or the Index 2XL. All autosequences stored by initialization, customization,
or remote download will only work with the instrument for which they were
designed.
To configure your medTester 5000C, begin by accessing the custom,
autosequence, SPO2 menu as described above (Customizing SPO2
Autosequences). This time, select F4 TYPE.
The display shows which SPO2 Simulator is currently selected.
Then select F1 INDEXL or F2 CARDIOSAT or F3 OXITEST+. Then
press F5STORE to store the selection. Now, all customized autosequences
will be for the designated SPO2 Simulator. After Changing the SPO2
Simulator, you must reinitialize your SPO2 autosequences so that they are
compatible with the change. See “Reinitialize Factory Default Settings”( page
230).
Making an Autosequence
Begin by accessing the custom, autosequence, SPO2 menu as described
above. This time, select:
MAKE
F1
Next select the name of the autosequence you want to customize or create.
Follow the following steps, providing the desired information:
1.
Name—This is the name you want to give to the autosequence,
with a maximum of six characters.
2.
MANUFACTURER: (CardioSat 100 and Index 2XL only) –
Select the manufacturer of the pulse oximeter to be tested with
F1
DEC
and
F2 INC
. Choose from the manufacturers in the table
appropriate for the desired tester below.
C
ARDIO
S
AT
100