Fluke Biomedical medTester 5000C User Manual
Page 55

Biomedical Test System
Manual Tests
4
4-9
Power Cord Resistance
See Figure 4-9, Power Cord Resistance Test.
1.
Plug the equipment under test into the
TEST RECEPTACLE
at the
top of the medTester 5000C.
2.
Connect the dual banana-plug end of a Kelvin cable to the two
red external meter posts.
3.
Connect the alligator-clip end of the Kelvin cable to a grounded
point on the case of the equipment under test.
4.
Enter menu commands:
OHMS
F4
MANUAL
F2
PWRCORD
F1
The
CURRENT SOURCE
lamp lights up to indicate that the
connection is made. The display reads the resistance (in OHMS)
from the case to the power cord ground pin.
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)