Reinitialize factory default settings – Fluke Biomedical medTester 5000C User Manual
Page 269

Biomedical Test System
Transcutaneous Pacemaker (Pacer) Module
16
16-13
Pacer Autosequence Number: A23
Autosequence Name: MDE3PM
Load: Adapter Module TQA-6, External 600 Ohms (Note: In Step 11 substitute 1000
Ohms.)
Step Number
Current
(in milliamperes)
Rate
(in PPM, pulses per minute)
1 60
40
2 60
60
3 60
70
4 60
75
5 60
80
6 60
90
7 60
100
8 60
120
9 60
140
10 60
160
11 120
60
12 90
100
13 120
100
Pulse Tests:
• Current Limit: ± 20%
• Rate Limit: ± 5 PPM
Demand Test: 80 PPM
• Underdrive: 30 BPM
• Ovedrive: 160 BPM
Async Test: 80 PPM
• Overdrive: 160 BPM
Refractory Period Test: No
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)