Leak test, Pop off test – Fluke Biomedical medTester 5000C User Manual
Page 279

Biomedical Test System
Non-Invasive Blood Pressure (NIBP) Module
17
17-5
Pretest Device Prompts
There are eight prompts for information that can appear in the autosequence 
test record. After selecting the desired autosequence, the medTester 5000C 
prompts you to enter device prompts. If you have some or all prompts turned 
off, the autosequence still gives you the opportunity to enter device 
information with the message: 
1.
OP CODE—This is the operator code for the person doing the 
testing. This code can consist of a maximum of three characters 
which can be entered from the medTester keyboard. An example 
of an operator code could be a person’s initials or employee 
number. The previous operator’s code, if any, will appear. If the 
current operator has not changed from previous use, press the 
ENT
key.
Sentinel users:
Substitute TECH CODE. This is
required for closing work orders.
2.
TYPE—Enter the device name, code, or other descriptor for the 
equipment under test. This entry can be a maximum of 16 
characters. 
3.
MANF—Enter the name of the device manufacturer, 16 
characters maximum. 
4.
LOC—Type in the location of the device, 16 characters 
maximum. The previous location, if any, will appear. If this occurs 
and is the desired location, press the 
ENT
key.
5.
MODEL—Enter the device model number, 16 characters 
maximum. 
6.
SERIAL #—Type the device serial number, again 16 characters 
maximum. 
7.
CONTROL #—Enter a control number with a maximum of 16 
characters. 
Sentinel users:
Use the control number found in the
equipment record.
8.
PHYSICAL INSPECTION—You have two physical inspection 
lines, each of which can consist of a maximum of 40 characters. 
After entering data for the first prompt, the second prompt 
appears. 
Press F5 for all device info prompts