Comments, Next test due date, User time – Fluke Biomedical medTester 5000C User Manual
Page 270: User time is required, Print test tag, Store or print test record

medTester 5000C
Operators Manual
16-14
Pacer Autosequence Number: A24
Autosequence Name: ZOLLPM
Load: Adapter Module TQA-2, External 1000 Ohms
Step Number
Current
(in milliamperes)
Rate
(in PPM, pulses per minute)
1 80
180
2 80
160
3 80
140
4 80
120
5 80
100
6 80
80
7 80
60
8 80
40
9 140
80
10 100
80
11 60
80
12 20
80
Pulse Tests:
• Current Limit: ± 20%
• Rate Limit: ± 5 PPM
Demand Test: 80 PPM
• Underdrive: 30 BPM
• Ovedrive: 160 BPM
Async Test: No
Refractory Period Test: No
Posttest Prompts
When all tests are completed, the medTester 5000C prompts you for:
1.
COMMENTS
2.
NEXT TEST DUE DATE
3.
USER TIME
EMS/CMMS system users:
User time is required
4.
PRINT TEST TAG
5.
STORE or PRINT TEST RECORD
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)