Serial port expansion, Keyboard interface, Port configuration – Fluke Biomedical medTester 5000C User Manual
Page 194

medTester 5000C
Operators Manual
12-16
WAVE COMMANDS
Note: The following commands turn on the specified wave. The wave continues to run until it
is changed by another wave command, or it will be turned off and disconnected by any non-
wave command or the EXIT command.
All wave commands return OK.
Command
Explanation
W000 Zero
output.
W001
2 Hz square wave.
W002
4 second pulse.
W003
10 Hz sine wave.
W004
40 Hz sine wave.
W005
60 Hz sine wave.
W006
100 Hz sine wave.
W007
1 kHz square wave.
W008
2 Hz triangle wave.
W009
60 Hz common mode rejection test.
W010
Normal Sinus ECG at 30 BPM.
W011
Normal Sinus ECG at 60 BPM.
W012
Normal Sinus ECG at 120 BPM.
W013
Normal Sinus ECG at 240 BPM.
W014
0.5 Hz sine wave.
W015
Atrial fibrillation 1, course.
W016
2nd degree AV block, type 1, Wenckebach.
W017
2nd degree AV block, type 2.
W018
Missed beat at 80 BPM.
W019
Missed beat at 120 BPM.
W020
PVC type 1, left.
W021
PVC type 2, right.
W022 Multifocal
PVCs.
W023
PVC 1, R on T.
W024
A pair or couplet of PVCs.
W025
Run of 5 PVCs.
W026
Run of 11 PVCs.
W027
Right bundle branch block.
W028
Ventricular tachycardia 180 BPM.
W029 Ventricular
fibrillation.
W030
Asystole 2, flat.