Print your autosequence, Reinitialize factory default settings – Fluke Biomedical medTester 5000C User Manual
Page 284

medTester 5000C
Operators Manual
17-10
NIBP Autosequence Number: A80
Autosequence Name: ADULT1
Step Number
Blood Pressure
Heart Rate in BPM
(beats per minute)
Envelope Gain
Envelope Shift
1
120/80 (90) ADULT
80
100%
+0
2
80/50 (62) ADULT
45
100%
+0
3
150/100 (115) ADULT
125
100%
+0
NIBP Test Limits:
• NIBP: ± 20%
• Heart Rate: ± 4 BPM
Pressure Test: No
Leak Test: No
Pop Off Test: No
NIBP Autosequence Number: A81
Autosequence Name: ADULT2
Step Number
Blood Pressure
Heart Rate in BPM
(beats per minute)
Envelope Gain
Envelope Shift
1
120/80 (90) ADULT
80
100%
+0
2
80/50 (62) ADULT
45
100%
+0
3
150/100 (115) ADULT
125
100%
+0
NIBP Test Limits:
• NIBP: ± 20%
• Heart Rate: ± 4 BPM
Pressure Test: Yes
Leak Test: Yes
Pop Off Test: Yes
NIBP Autosequence Number: A82
Autosequence Name: ADULT3
Step Number
Blood Pressure
Heart Rate in BPM
(beats per minute)
Envelope Gain
Envelope Shift
1
120/80 (90) ADULT
80
100%
+0
2
80/50 (62) ADULT
80
100%
+0
3
150/100 (115) ADULT
80
100%
+0
4
120/80 (90) ADULT
45
100%
+0
5
120/80 (90) ADULT
125
100%
+0
6
200/150 (165) ADULT
200
100%
+0
7
120/80 (90) AFIB
N/A
N/A
N/A
8
120/80 (90) PAC
N/A
N/A
N/A
NIBP Test Limits:
• NIBP: ± 20%
• Heart Rate: ± 4 BPM
Pressure Test: Yes
Leak Test: Yes
Pop Off Test: Yes