Iv pump module, Iv pump autosequences, Configuring for your iv pump analyzer – Fluke Biomedical medTester 5000C User Manual
Page 135

Biomedical Test System
Defibrillator Module
9
9-11
The table below supplies specific information about the tests for each
autosequence.
Autosequence
Name
Energy
Step High
Limit
(%)
Energy
Step Low
Limit
(%)
Max
Energy
High
Limit
(Joules)
Max
Energy
Low Limit
(Joules)
Paddle
Leakage
and/or
Cardioversion
Energy Steps
(Joules)
LP4
10
10
480
418
Cardioversion
10, 25, 50, 100,
200, 300, 400, 450
LP5
10
10
396
324
Cardioversion
20, 50, 100, 200,
300, 360
LP6
10
10
428
372
Both
5, 10, 20, 30, 50,
100, 200, 300, 400
LP6S
10
10
385
335
Both
5, 10, 29, 30, 50,
100, 150, 200, 300,
360
LP7
10
10
385
335
Both
5, 10, 20, 30, 50,
100, 200, 300, 360
LP8
10
10
385
335
Both
2, 5, 9, 10, 20, 30,
50, 100, 150, 200,
300, 360
M/D3
10
10
440
360
Both
5, 10, 20, 40, 60,
100, 150, 200, 250,
300, 360
78660A
10
10
400
330
Both
5, 10, 20, 40, 60,
100, 150, 200, 250,
300, 360
NK7000
10
10
396
324
Both
3, 5, 10, 20, 30, 50,
70, 100, 150, 200,
300, 360
AOLOWN
15
15
414
306
Both
5, 10, 20, 30, 50,
100, 150, 200, 300,
360
M/D4
20
20
384
256
Both
1, 8, 40, 80, 160,
240, 320
M/D2J
10
10
506
404
Both
5, 10, 20, 40, 75,
100, 200, 300, 400,
460
78670A
15
15
414
306
Both
5, 10, 20, 30, 50,
70, 100, 150, 200,
300, 360
43100A
15
15
414
306
Both
2, 3, 5, 7, 10, 30,
50, 70, 100, 150,
200, 300, 360
ZOLL
10
10
396
324
Both
2, 3, 5, 7, 10, 30,
50, 100, 150, 200,
300, 360
LP10
10
10
396
324
Cardioversion
5, 10, 20, 50, 100,
200, 300, 360
3 blanks
(user-defined)