To set up and connect a label printer, Selecting a label printer – Fluke Biomedical medTester 5000C User Manual
Page 80

medTester 5000C
Operators Manual
5-10
ECG Performance Waveforms
If you have the Waveform/Extended Testing module installed on your
medTester 5000C, and you have performed a lead test, this message appears:
This is a prompt to output an autosequence of performance waves to a
patient monitor from the patient leads attached to the medTester 5000C.
Waveforms are output in the order described in the table below for various
periods of time.
Waveform Name
Test Duration
(In Seconds)
2 Hz Square
Continuous
4-second Pulse
8
10 Hz Sine
4
40 Hz Sine
4
60 Hz Sine
60
100 Hz Sine
4
1 kHz Square
4
2 Hz Triangle
Continuous
CMRR 60 Hz
Continuous
ECG 60 BPM
Continuous
ECG 30 BPM
30
ECG 60 BPM
30
ECG 120 BPM
30
ECG 240 BPM
30
Note
Tests with a predefined duration automatically end. However,
you can choose to terminate them at any time by pressing
NEXT
.
Those marked continuous can be terminated by pressing
NEXT
.
OUTPUT PERFORMANCE WAVES?