For the oxitest plus – Fluke Biomedical medTester 5000C User Manual
Page 241

Biomedical Test System
SPO2 Module
15
15-13
For the Oxitest PLUS:
SPO2 Autosequence Number: A70
Autosequence Name: ADULT1
Settling Time: 25 Beats
Heart Rate Limit:
± 4 BPM (beats per minute)
Step Number
SPO2
( in percent)
SPO2 Limit
(in percent)
Heart Rate
(in BPM)
Amplitude
(in percent)
Artifact
1 97 ±4
75 100 Off
2 97 ± 4
45 100 Off
3 97 ± 4
180 100 Off
4 80 ± 4
45 100 Off
5 90 ± 4
120 100 Off
6 80 ± 4
180 100 Off
SPO2 Autosequence Number: A71
Autosequence Name: ADULT2
Settling Time: 25 Beats
Heart Rate Limit:
± 4 BPM (beats per minute)
Step
Number
SPO2
( in percent)
SPO2 Limit
(in percent)
Heart Rate
(in BPM)
Amplitude
(in percent)
Artifact
1 97 ±4
75 100 Off
2 97 ± 4
45 100 Off
3 97 ± 4
180 100 Off
4 80 ± 4
45 100 Off
5 90 ± 4
120 100 Off
6 80 ± 4
180 100 Off
Starting Alarm Value for SPO2: 93%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 100%
SPO2 Alarm Test: Hi–97% Lo–80%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
Probe Test