Leakage current measurements, Equipment current measurement – Fluke Biomedical medTester 5000C User Manual
Page 75

Biomedical Test System
Autosequences
5
5-5
SAFETY AUTOSEQUENCES
TEMPMON Temperature
Monitor
TOOLBOX
F2
TEMPMON
F4
A18
HUMMON Humidity
Monitor
TOOLBOX
F2
HUMMON
F5
A19
Autosequence Device Prompts 
After you select an autosequence, medTester prompts you to enter the 
following information for the device that you want to test. Entry information 
can be any combination of alpha or numeric characters. After each entry press 
the 
ENT
(Enter) key.
1.
OP CODE—This is the operating code for the person doing the 
testing. This code can consist of a maximum of three characters 
which can be entered from the medTester keyboard. An example 
of an operator code could be a person's initials or employee 
number. The previous operator’s code, if any, will appear. If the 
current operator has not changed from previous use, press the 
ENT
key.
2.
TYPE—Enter the device name, code, or other descriptor for the 
equipment under test. This entry can be a maximum of 16 
characters. 
3.
MANF—Enter the name of the device manufacturer, 16 
characters maximum. 
4.
LOC—Type in the location of the device, 16 characters 
maximum. The previous location, if any, will appear. If this occurs 
and is the desired location, press the 
ENT
key.
5.
MODEL—Enter the device model number, 16 characters 
maximum. 
6.
SERIAL #—Type the device serial number, again 16 characters 
maximum. 
7.
CONTROL #—Enter a control number with a maximum of 16 
characters. 
8.
PHYSICAL INSPECTION—You have two physical inspection 
lines, each of which can consist of a maximum of 40 characters. 
After entering data for the first prompt, the second prompt 
appears. 