End-of-test prompts, Toolbox monitors, Temperature monitor – Fluke Biomedical medTester 5000C User Manual
Page 85: Optional monitoring autosequences

Biomedical Test System
Autosequences
5
5-15
1.
From the menu, press
F1
(
CONFIG
) as before to configure your
test tag (test tag lines on/off), or press
F2
(
PRINTER
) to select a
label printer.
2.
From the (
PRINTER
) menu press
F1
(
TLP
) to select Fluke
Biomedical TLP printers (TLP-1 or TLP-2), or
F2
(
BRADY
) to
select the Brady printer. The selection is displayed.
3.
Press
F5
(
STORE
) to store the selection in nonvolatile
EEPROM memory.
Note
When using the Brady printer, it is not necessary to set the
number of lines on the label with the
UTIL / PRTPAGE / TAGS
menu path. However, you still need to select the fields you want
printed on the label through the
CUSTOM
/
TAG
/
CONFIG
menu
path.
To configure a medTester serial port for test tag printing, see
Local Output for Records and Tags in Chapter 12, Remote
Operation. To configure one of the Wedge adapter COM2 ports
for test tag printing, see Setting Output Ports in Chapter 13, The
Wedge Adapter
Optional Monitoring Autosequences
Two optional autosequences are available in the Waveform/Extended Testing
Module. Both are designed as monitors.
•
Line monitor voltage test—periodically measures line voltage, and
•
Environmental test—which measures ground potential and
resistance between a common ground point and up to 99 other
ground points in a room.
MONITORING AUTOSEQUENCES
Autosequence
Name
Function
Menu Command
From MENU 2
Remote
Command
LINEMON Line
voltage
monitor.
LINEMON
F4
AUTO
F1
A16
ENVMON Ground
potential
and resistance.
ENVMON
F5
AUTO
F1
A17