Error responses, Local output for records and tags – Fluke Biomedical medTester 5000C User Manual
Page 182

medTester 5000C 
Operators Manual 
12-4
You can connect the optional RS-232 keyboard or laser bar code gun to either 
serial port. Input from this keyboard appears on the medTester display 
identical to the appearance of input displayed from the medTester 5000C 
keyboard. Hexadecimal codes for mimicking keystrokes are as follows. Any 
serial connecting device can send the codes with the same results: 
Hexadecimal Number to Enter
Corresponding medTester Key
80 F1 
81 F2 
82 F3 
83 F4 
84 F5 
11 Cursor 
UP
14 Cursor
DOWN
12 Cursor
LEFT
13 Cursor
RIGHT
1B ESCAPE 
0D—Carriage return (CR), or 
0A—Line feed (LF) 
ENTER
08 BACKSPACE 
B0 INSERT 
7F DELETE 
10 (Control-P) 
All ASCII alphanumeric and punctuation codes from 20 (space) to 7E (~). 
Note: In the case where strings are entered rapidly by computer, they may be terminated by both CR and 
LF.
Local Output for Records and Tags
In local mode, you can output test records, instrument settings, and test tags 
via COM1, COM2, or the printer port. To configure output for records or 
tags, once again use the utility menu: 
OUTPUT
F2
UTIL
F5
Menu choices displayed are:
RECORDS
F1
and
TAGS
F2