11
11-5
As you can see, some checkitems appear in the medTester as prompts to the 
operator to do something, for example, a physical inspection of the EUT or a 
check of some function. Other checkitems are autosequences. Autosequence 
46 is a customized (user defined) autosequence that can execute a variety of 
performance tests on this defibrillator. For more information about 
defibrillator autosequences, see Chapter 9, Defibrillator Module. To learn more 
about customizing autosequences, see Chapter 6, Customize Your 
medTester 5000C. 
This checklist could be downloaded to the medTester with a list of control 
numbers. Control numbers identify unique pieces of equipment. Each control 
number has a checklist assigned to it. When you perform the checklist on a 
control number-referenced piece of equipment, the test record that is 
generated includes the control number. If you test a defibrillator of this model 
type and it has no control number, the medTester can prompt you to assign 
one. This lets you bring new equipment into inventory as you test. Of course, 
you still have the option of making an equipment record and assigning a 
control number prior to testing it with the checklist. 
In this example the last checkitem in the checklist prompts the operator to 
write out a preventive maintenance tag that is placed on the EUT after the 
checklist runs. The medTester 5000C also has an automated tag generation 
capability. This requires the connection of an optional test label printer. See 
the Test Tag Printer section in Chapter 5. 