Biomedical Test System
ESU Module
14
14-13
Autosequence Number: A63
Autosequence Name: VLF4BF
Step and Test
Type
Load
ESU Mode
Output
Power
Setting
Expected Power
or
Electrode
Limits
1
Generator Output
300 Ohms
Monocoag
Spray
50
50 W
± 20%
2
Generator Output
300 Ohms
Monocoag
Spray
120
120 W
± 20%
3
Generator Output
300 Ohms
Monocoag
Soft
60
60 W
± 20%
4
Generator Output
300 Ohms
Monocoag
Low V
99
99 W
± 20%
5
Generator Output
300 Ohms
Monopolar
Pure Cut
50
50 W
± 20%
6
Generator Output
300 Ohms
Monopolar
Pure Cut
300
300 W
± 20%
7
Generator Output
300 Ohms
Monopolar
Blend 1
250
250 W
± 20%
8
Generator Output
300 Ohms
Monopolar
Blend 2
50
50 W
± 20%
9
Generator Output
300 Ohms
Monopolar
Blend 2
200
200 W
± 20%
10
Generator Output
300 Ohms
Monopolar
Blend 3
200
200 W
± 20%
11
Generator Output
100 Ohms
Bipolar
Standard
70
70 W
± 20%
12
Generator Output
100 Ohms
Bipolar
Precise
70
70 W
± 20%
13
Auxiliary Return
Fault Test
N/A Cut
Side N/A 300
N/A
N/A
14
Auxiliary Return
Fault Test
N/A Coag
Side N/A
300
N/A
N/A
15
Auxiliary CQM
Test
N/A N/A N/A N/A N/A
N/A