Fluke Biomedical medTester 5000C User Manual
Page 208

medTester 5000C
Operators Manual
13-12
Note
Autosequences always communicate at a fixed baud rate;
therefore, you don't need to set the rate for any port used only
for autosequences.
To set a Wedge port baud rate:
1.
From the medTester 5000C main menu,
MENU 1
, press:
UTIL
F5
2.
From the utility menu, press:
BAUD
F1
to see this display:
Pressing the medTester’s right arrow lets you access the remaining port:
3.
Press the function key that corresponds to the port you want, for
example COM2-1.
4.
Press
F4
to change the baud rate value that appears in the display
until you reach the value you want:
COM PORT BAUD RATES
COM2-4
F5
COM2-3
F4
COM1
F1
COM2-2
F3
COM2-1
F2
COM PORT BAUD RATES
COM2-8
F4
COM2-5
F1
COM2-7
F3
COM2-6
F2
SET COM2-1 BAUD RATE:
9600
BAUD
F4
STORE
F5
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)