Spo2 alarm test, Rate alarm test, Asystole test – Fluke Biomedical medTester 5000C User Manual
Page 235: Amplitude test

Biomedical Test System
SPO2 Module
15
15-7
Note
The Index 2XL must have been manufactured after 5/21/04 or
modified by Fluke Biomedical (ECO 4163) to communicate
through the Wedge Port Expander.
Note
Refer to Chapter 1, General Information – Accessories, or the current
Fluke Biomedical Price List for the correct serial cable part
number needed.
From the medTester 5000C SPO2 menu, select the autosequence name for
the autosequence you wish to run. The name and number of the selected
autosequence appears on the display:
If a safety autosequence has executed since power up of the medTester, you
can have the medTester use the previously input prompt information for the
current SPO2 autosequence, from this prompt:
To accept previously input prompt information, press
YES
, and go to Test
Sequence below. If you choose
NO
, go to the prompt instructions below.
Press F5 for all device info prompts
USE PREVIOUS STARTING INFORMATION?
YES
F1
NO
F2