Biomedical Test System
IV Pump Module
10
10-21
IPT-MC AUTOSEQUENCE TESTS
Nominal
Flow Rate
(ml per
hour)
High
Limit
(ml per
hour)
Low
Limit
(ml per
hour)
End Type
(manual,
volume,
or timed)
Ending Quantity
If Applicable
(ml for volume or
hours:minutes
for timed)
Autosequence Name (Test Type): PUMP 1
Keyboard Shortcut Command: A50
60.0 66.0 54.0
Volume
5.0
120.0 132.0 108.0
Volume
10.0
240 264.0
216.0
Volume 20.0
Autosequence Name (Test Type): PUMP 2
Keyboard Shortcut Command: A51
30.0 35.0 25.0
Volume
3.0
60.0 66.0 54.0
Volume
5.0
120.0 132.0 108.0
Volume
10.0
120.0 132.0 108.0
Volume
10.0
Autosequence Name (Test Type): LOW
Keyboard Shortcut Command: A52
10.0 11.0 9.0 Timed 00:05
Autosequence Name (Test Type): HIGH
Keyboard Shortcut Command: A53
480.0 504.0 456.0 Timed
00:05
Autosequence Name (Test Type): EVAL
Keyboard Shortcut Command: A54
5.0 5.9 4.1
Volume 1.5
10.0 11.0 9.0
Volume
2.5
20.0 22.0 18.0
Volume
2.5
30.0 33.0 27.0
Volume
2.5
60.0 66.0 54.0
Volume
5.0
120.0 132.0 108.0
Volume
10.0
240.0 264.0 216.0
Volume
30.0
480.0 528.0 432.0
Volume
60.0
240.0 264.0 216.0
Volume
30.0
480.0 528.0 432.0
Volume
60.0