Storing an autosequence record – Fluke Biomedical medTester 5000C User Manual
Page 78

medTester 5000C 
Operators Manual 
5-8
System Line Voltage
Measurements included are:
•
Hot to neutral,
•
Neutral to ground, and
•
Hot to ground.
Power Cord Resistance
After taking measurements of line voltage, the autosequence moves to the 
power cord resistance measurement. The medTester offers you the option of 
skipping the test. You could skip a resistance measurement, for example, with 
double-insulated equipment having no exposed metal surfaces. If you choose 
to skip this test, it is recorded as being skipped. To skip the measurement, 
from the menu select: 
SKIP
F4
You can proceed with the resistance test, but first notice the displayed reading 
and 
CURRENT SOURCE
annunciator. If any of the following appear on the
medTester display, it means a condition exists that prohibits an accurate 
measurement: 
•
# (pound) sign—Indicates an out-of-limit condition based upon the 
resistance limit set for the autosequence you are running. Limits are 
programmable. 
•
x.xxx—Indicates that a current source is not connected.
•
Flashing reading—Indicates that the displayed reading is out of 
range, that is >2 ohms. 
•
CURRENT SOURCE annunciator not lit—Indicates lack of a good 
connection. 
Correct the cause of the condition before continuing. For example, reconnect 
the Kelvin cable to an exposed grounded surface on the EUT. If none of 
these conditions exist and a measurement appears in the display, press: 
READY
F5